Prosthesis design influences segmental contribution to total cervical motion after cervical disc arthroplasty.

Icon for SpringerRelated Articles

Prosthesis design influences segmental contribution to total cervical motion after cervical disc arthroplasty.

Eur Spine J. 2019 Jul 15;:

Authors: Patwardhan AG, Havey RM

Abstract
INTRODUCTION: We investigated a new metric for assessing the quality of motion of the cervical segments over the arc of extension-to-flexion motion after cervical disc arthroplasty (CDA). We quantified: (1) the amount of motion contributed by individual spinal segments to the total cervical spine motion, termed segmental motion fraction, and its variation throughout the arc of extension-to-flexion motion and (2) how cervical disc arthroplasty using two distinct prosthesis designs may influence the segmental motion contributions.
MATERIALS AND METHODS: We tested 16 human C3-T1 spine specimens under physiologic loads; first intact, after CDA at C5-C6, and then at C5-C6 and C6-C7. The M6-C (Orthofix, USA) and Mobi-C (Zimmer, USA) disc prostheses were used in eight specimens each.
RESULTS AND CONCLUSIONS: The designs of the cervical disc prostheses tested significantly influenced the variation in segmental motion fraction as the spine underwent motion between the endpoints of extension and flexion. While the mean segmental motion contribution to the total cervical motion was not influenced by prosthesis design, the way the motion took place between the extension and flexion endpoints was significantly influenced. The M6-C artificial disc restored physiologic motion quality such that implanted segments continued to function in harmony with other segments of the cervical spine as measured before arthroplasty. Conversely, the Mobi-C prosthesis, while maintaining average motion contributions similar to the pre-implantation values, demonstrated large deviations in motion contribution over the extension-to-flexion arc motion in ten of 16 implanted segments. Such non-physiologic implant kinematics could cause excessive prosthesis wear and motion and stress shielding at adjacent segments. These slides can be retrieved under Electronic Supplementary Material.

PMID: 31309331 [PubMed – as supplied by publisher]

Prosthesis design influences segmental contribution to total cervical motion after cervical disc arthroplasty | Sciatica specialist

What our patients say ...

Consultant Spine Surgeon
Consultant Spine Surgeon
Consultant Spine Surgeon

This surgical technique consists of a percutaneous approach for the treatment of small to medium size hernias of the intervertebral disc by laser energy. The main objective is to reduce the intradiscal pressure in the nucleus pulposus

Laser Disc Surgery can be performed under local anaesthetic as a day case at our centre on the prestigious Harley Street.
What is London spine unit and How it Works

The London Spine Unit was established in 2005 and has successfully treated over 5000 patients. All conditions are treated.

treatment of all spinal disorders

The London Spine Unit specialises in Minimally Invasive Treatments allowing rapid recovery and return to normal function

Trusted by patients worldwide

The London Spine Unit provides the highest quality care to all patients and has VIP services for those seeking exceptional services

If you have any emergency Doctor’s need, simply call our 24 hour emergency

Your personal case manager will ensure that you receive the best possible care.

Call Now 

+44 844 589 2020
+44 203 973 8810