STUDY DESIGN: Finite element and in vitro study. OBJECTIVE: Finite element calculations to delineate a dynamic fixator and confirmation with an in vitro experiment. SUMMARY AND BACKGROUND DATA: In the last few years, there was a paradigm shift from rigid to dynamic fixation of spinal segments. However, some so-called dynamic implants like the Dynesys performed still stiffer than anticipated. The aim of this study was to optimize a dynamic stabilization system. METHODS: The development steps of this implant design can be summarized in a development loop. First, a finite element model of an intact human L4-L5 segment was used to delineate implant stiffness parameters for the implant, in consideration of clinical concerns and safety aspects. These data were used in a second step, leading to the final implant design. This development process was completed with an appropriate in vitro experiment. The optimal axial and bending stiffness were computed to reduce the spinal motion by 30%. For the validation process, in vitro tests were performed on 6 human lumbar spinal segments L2-L3 with a median age of 52. The model and the specimens were loaded with pure unconstrained moments of 7.5 Nm in flexion, extension, lateral bending, and axial rotation. RESULTS: This study demonstrated the advantages of employing a finite element model for the implant parameter delineation. It was possible to prospectively outline the needed stiffness parameters for a desired spinal range of motion achievement. CONCLUSION: In summary, FEM may accelerate the development and the realization of a new implant design

Keywords : adverse effects,anatomy & histology,Biomechanical Phenomena,Biomedical Engineering,Cadaver,Equipment Design,Equipment Safety,etiology,Finite Element Analysis,Germany,Humans,instrumentation,Internal Fixators,Lumbar Vertebrae,Materials Testing,methods,Middle Aged,Motion,physiology,physiopathology,Postoperative Complications,prevention & control,Prospective Studies,Prostheses and Implants,Range of Motion,Articular,Rotation,Safety,Spinal Fusion,standards,surgery,Universities,Weight-Bearing,, Design,Delineation,Subsequent, massage for back pain near me

Date of Publication : 2009 Feb 1

Authors : Wilke HJ;Heuer F;Schmidt H;

Organisation : Institute of Orthopaedic Research and Biomechanics, University of Ulm, Ulm, Germany. hans-joachim.wilke@uni-ulm.de

Journal of Publication : Spine (Phila Pa 1976 )

Pubmed Link : https://www.ncbi.nlm.nih.gov/pubmed/19179920

The London Spine Unit : Harley Street UK. Specialists in Cutting Edge Technologies for Spinal Surgery

Make an Appointment 

Trustpilot Reviews
Doctify Reviews
Top Doctor Reviews

Prospective design delineation and subsequent in vitro evaluation of a new posterior dynamic stabilization system | Home clinic tens machine

What our patients say ...

Consultant Spine Surgeon
Consultant Spine Surgeon
Consultant Spine Surgeon

This surgical technique consists of a percutaneous approach for the treatment of small to medium size hernias of the intervertebral disc by laser energy. The main objective is to reduce the intradiscal pressure in the nucleus pulposus

Laser Disc Surgery can be performed under local anaesthetic as a day case at our centre on the prestigious Harley Street.
What is London spine unit and How it Works

The London Spine Unit was established in 2005 and has successfully treated over 5000 patients. All conditions are treated.

treatment of all spinal disorders

The London Spine Unit specialises in Minimally Invasive Treatments allowing rapid recovery and return to normal function

Trusted by patients worldwide

The London Spine Unit provides the highest quality care to all patients and has VIP services for those seeking exceptional services

If you have any emergency Doctor’s need, simply call our 24 hour emergency

Your personal case manager will ensure that you receive the best possible care.

Call Now 

+44 844 589 2020
+44 203 973 8810