The London Spine Unit : most advanced treatment hospital in the world
Published article
CONCLUSIONS: On a group level, both components of the Prodisc-C Vivo cervical disc arthroplasty remained stable over time and below the clinical threshold of 1 mm. Individual outliers for translation were not clinically relevant and probably related to settling of the components into the vertebral endplates. RSA allowed us to perform a preliminary but accurate study on the micromotion of a new cervical disc replacement in a small sample size, without putting large numbers of patients at risk.Cervical Disc Replacement Expert. Best Spinal Surgeon UK
The London Spine Unit is a specialist clinic for Cervical Disc Replacement as Day Surgery.
Abstract
Study design: Prospective cohort study.
Objective: To investigate the primary stability of the Prodisc-C Vivo cervical disc arthroplasty with regard to the adjacent cervical vertebrae using radiostereometric analysis (RSA), and to monitor its clinical performance.
Methods: Sixteen patients with degenerative cervical disc disease were included. RSA radiographs were obtained at the first postoperative day, at 6 weeks, 3 months, and 6 months postoperatively. Migration (translation ) of the superior and inferior implant components were measured with model-based RSA, expressed along the 3 orthogonal axes, and calculated as total translation. Clinical outcomes were Neck Disability Index, numeric rating scales for neck and arm pain, Likert-type scales for satisfaction, and adverse events. Range of motion was reported as C2-C7 flexion-extension mobility (ROM).
Results: At final follow-up, no significant increase over time in median total translation was found. One inferior and 3 superior components subsided but were asymptomatic. ROM remained stable and clinical outcomes improved over time. Although 3 patients were unsatisfied and 3 adverse events occurred, this was not related to translation of the components.
Conclusions: On a group level, both components of the Prodisc-C Vivo cervical disc arthroplasty remained stable over time and below the clinical threshold of 1 mm. Individual outliers for translation were not clinically relevant and probably related to settling of the components into the vertebral endplates. RSA allowed us to perform a preliminary but accurate study on the micromotion of a new cervical disc replacement in a small sample size, without putting large numbers of patients at risk.
Keywords: arthroplasty; implants; outcome; radiostereometric analysis; spine.
The London Spine Unit : most advanced treatment hospital in the world
Read the original publication from Pubmed :