STUDY DESIGN: Randomized, controlled, multicenter, investigational device exemption trial. OBJECTIVE: To investigate the safety and effectiveness of the first two-piece, metal-on-metal lumbar disc prosthesis for treating patients with single-level degenerative disc disease. SUMMARY OF BACKGROUND DATA: For patients with degenerative disc disease unresponsive to conservative measures, lumbar disc arthroplasty provides an alternative to fusion designed to relieve persistent discogenic pain and maintain motion. METHODS: After 2:1 randomization, 577 patients were treated in either the investigational group (405), receiving lumbar disc arthroplasty, or the control group (172), receiving anterior lumbar interbody fusion. Patients were evaluated preoperatively, at surgery/discharge, and at 1.5, 3, 6, 12, and 24 months after surgery. The primary study endpoint was overall success, a composite measure of safety and effectiveness as recommended by the Food and Drug Administration and defined in the protocol. RESULTS: Both treatment groups demonstrated significant improvements compared with preoperative status. The investigational group had statistically superior outcomes (P < 0.05) at all postoperative evaluations in Oswestry Disability Index, back pain, and Short Form-36 Physical Component Summary scores as well as patient satisfaction. Investigational patients had longer surgical times (P < 0.001) and greater blood loss (P < 0.001) than did control patients; however, hospitalization stays were similar for both groups. Investigational patients had fewer implant or implant/surgical procedure-related adverse events (P < 0.001). Return-to-work intervals were reduced for investigational patients. Disc height and segmental angular motion were maintained throughout the study in the investigational group. In the investigational group, overall success superiority was found when compared to the control group as defined by the Food and Drug Administration Investigational Device Exemption protocol. CONCLUSION: The investigational group consistently demonstrated statistical superiority versus fusion on key clinical outcomes including improved physical function, reduced pain, and earlier return to work Keywords : Adolescent,Adult,adverse effects,Aged,Arthroplasty,Back Pain,etiology,Female,Follow-Up Studies,Humans,Intervertebral Disc,Intervertebral Disc Degeneration,Lumbar Vertebrae,Male,methods,Middle Aged,Motion,Pain,Pain Measurement,Pain,Postoperative,Patient Satisfaction,physiopathology,Prospective Studies,Prostheses and Implants,Range of Motion,Articular,Spinal Fusion,surgery,Surveys and Questionnaires,Time Factors,Treatment Outcome,Young Adult,, Disc,Arthroplasty,With,Maverickâ„¢,Disc, do radio frequency treatments really work
Date of Publication : 2011 Dec 1
Authors : Gornet MF;Burkus JK;Dryer RF;Peloza JH;
Organisation : The Orthopedic Center of St. Louis, St. Louis, Missouri 63017, USA. mfgspine@gmail.com
Journal of Publication : Spine (Phila Pa 1976 )
Pubmed Link : https://www.ncbi.nlm.nih.gov/pubmed/21415812
The London Spine Unit : Harley Street UK. Specialists in Cutting Edge Technologies for Spinal Surgery