19 Harley St, London, W1G 9QJ, UK

Failure of resorbable plates and screws in an ovine model of anterior cervical discectomy and fusion

BACKGROUND CONTEXT: Containment plates are often placed anteriorly in anterior cervical discectomy and fusion (ACDF) to provide stability and prevent migration of the interbody device or autograft. The main advantage of a bioresorbable plate over the typical metallic plate is that it will resorb after fusion has occurred, thus mitigating any long-term instrumentation-related complications. Furthermore, the plates are radiolucent, allowing complete visualization of the fusion site and eliminating imaging artifact. PURPOSE: To evaluate radiographic fusion, mechanical success rates, and histologic characteristics of a bioresorbable containment plate and screws in a 3-month ovine model of ACDF. STUDY DESIGN: An in vivo prospective analysis of resorbable anterior cervical plates and screws for use in ACDF in an ovine model. METHODS: Six sheep underwent C2-C3 and C4-C5 discectomies. Fusions were performed using a polyetheretherketone cage filled with autograft bone. A polymeric plate (70/30 poly-dl-lactic acid), and four screws were placed over an intervertebral disc spacer at each of these two levels. After 3 months, the animals were euthanized and radiographically imaged. Radiographs were analyzed for fusion and instrumentation failures. Functional spinal units were harvested for histologic processing and evaluation. RESULTS: Radiographic fusion was noted in three of the 12 levels with no evidence of device failure at any of the levels. However, at necropsy, it was observed that six of the 12 specimens had either a broken screw or a cracked plate. These gross observations were confirmed within the histologic sections. Fusion was verified histologically at C2-C3 in three of the six sheep; none of the fusions were successful at C4-C5. Histologic analysis also found that the tissue surrounding the plate and disc spacer consisted of vascularized fibrous tissue with islands of active woven bone. Inflammatory cells were rarely observed. CONCLUSIONS: Although the bioresorbable plates and screws did not elicit an iatrogenic tissue response, a high percentage of them failed mechanically. This phenomenon was difficult to observe radiographically, as the radiolucent markers were not able to convey these instrumentation failures. Additionally, there was only a 25% fusion rate. These findings suggest that resorbable implant materials with the current biomechanical and chemical properties are inadequate for cervical fusion. The results of this study strongly suggest that radiographic outcomes alone may not be adequate and that gross or histologic methods should accompany radiographs in studies of bioresorbable materials in animal models

Keywords : Absorbable Implants,analysis,Animals,Bone Plates,Bone Screws,Cervical Vertebrae,complications,Diskectomy,Equipment Failure,instrumentation,Intervertebral Disc,methods,Sheep,Spinal Fusion,, Resorbable,Plates,Screws, repetitive strain injury symptoms in hand

Date of Publication : 2011 Sep

Authors : Lyons AS;Sherman BP;Puttlitz CM;Patel VV;Abjornson C;Turner AS;Seim HB;Burger EL;Lindley EM;

Organisation : Orthopaedic Bioengineering Research Laboratory, Department of Mechanical Engineering, Colorado State University, Fort Collins, CO 80523, USA

Journal of Publication : Spine J

Pubmed Link : https://www.ncbi.nlm.nih.gov/pubmed/21775217

The London Spine Unit : Harley Street UK. Specialists in Cutting Edge Technologies for Spinal Surgery

Make an Appointment 

Trustpilot Reviews
Doctify Reviews
Top Doctor Reviews

Failure of resorbable plates and screws in an ovine model of anterior cervical discectomy and fusion | Cervical disc replacement | Circle Health

What our patients say ...

Consultant Spine Surgeon
Consultant Spine Surgeon
Consultant Spine Surgeon

This surgical technique consists of a percutaneous approach for the treatment of small to medium size hernias of the intervertebral disc by laser energy. The main objective is to reduce the intradiscal pressure in the nucleus pulposus

Laser Disc Surgery can be performed under local anaesthetic as a day case at our centre on the prestigious Harley Street.
What is London spine unit and How it Works

The London Spine Unit was established in 2005 and has successfully treated over 5000 patients. All conditions are treated.

treatment of all spinal disorders

The London Spine Unit specialises in Minimally Invasive Treatments allowing rapid recovery and return to normal function

Trusted by patients worldwide

The London Spine Unit provides the highest quality care to all patients and has VIP services for those seeking exceptional services

If you have any emergency Doctor’s need, simply call our 24 hour emergency

Your personal case manager will ensure that you receive the best possible care.

Call Now 

+44 844 589 2020
+44 203 973 8810