This is the first report of M6-C failure after conversion of an adjacent arthroplasty to fusion. A growing number of reports surrounding the M6-C failure rate and mechanisms raise concern about the device’s durability and underscore the importance of routine clinical and radiographic surveillance for these patients.
Cervical Disc Arthroplasty Expert. Best Spinal Surgeon UK
The London Spine Unit is a specialist clinic for Cervical Disc Arthoplasty as Day Surgery.
Abstract
Case: We present a delayed failure of an M6-C cervical disc arthroplasty after conversion of a subjacent failed cervical disc arthroplasty to fusion. The annular component had failed, and the core had been ejected. Histology demonstrated a giant cell reaction to polyethylene debris, and tissue cultures were positive for Cutibacterium acnes.
This is the first report of M6-C failure after conversion of an adjacent arthroplasty to fusion. A growing number of reports surrounding the M6-C failure rate and mechanisms raise concern about the device’s durability and underscore the importance of routine clinical and radiographic surveillance for these patients.